Main content of this page

Anchor links to the different areas of information in this page:

You are here: Companies & Products. Exhibitors & Products 2011.

Product Category Search

Product Category Search

Show location in hall

AGD Biomedicals Pvt. Ltd.
1-2, Shakti Mistry Complex, J.B. Nagar Andheri (East)
400 059 Mumbai, India

Phone: +91 22 28231061
Fax: +91 22 28224297
http://www.agdbio.com

Manufacturers Manufacturers
Service Service
OEM OEM

Login

for Exhibitors
for Visitors

Interactive Plan

Product Information
Product Categories
Product
Information
Analyser appliances  
Photometers  
Electrolytes (without electrodes)  
Electrolyte analyses using ISE (ion selective electrodes) 1 Product
Reagents for electrolyte analyses  
Devices and systems for haemoglobin determination 1 Product
Units and systems for clinical chemistry  
Semi-automatic units for clinical chemistry 2 Products
Haematology cell counters  
Haematology reagents 1 Product
27.10.2011

Specifications

  • Method: Ion-selective electrodes
  • Measured parameters: Na, K, Cl
  • Sample type: Whole blood, Serum, Plasma, urine & body fluids
  • Sample volume: 130 µl
  • Sample presentation: Syringe, test tube or capillary
  • Measurement time: 30 seconds
  • Calibration: Automatic, manual & one point calibration before each sample
  • Washing mode: Automatic
  • Measuring range:
  • K: 0.5 - 15.00 mmol/L
  • Na: 30.00 - 200.00 mmol/L
  • Cl: 30.00 - 200.00 mmol/L
  • CV:
  • K: < 1.0%
  • Na: < 1.0%
  • Cl: < 1.0%
  • Urine analysis: 5 user-defined dilution ratios
  • QC programs: Trilevel (Low, Normal & High) with L-J plot
  • Search data: Date wise, sample type wise, patient ID wise
  • Data storage: 1000 samples
  • Error messages: Insufficient reagents/sample
  • Display size: 240 x 64 dots backlit LCD
  • Serial port: RS 232
  • Printer: 20 column thermal printer
  • Power supply: 110-220V +/- 5%, 50-60 Hz A.C.
  • Dimensions: 33 x 32 x 34 cms
  • Weight: Approx. 10 kg

 

Salient Features:

  • Simple operation.
  • Fast sample processing.
  • Flow through, long life, maintenance free electrodes.
  • Low sample volume.
  • Standby mode when instrument is not in use for 10 minutes.
  • Low cost per test.

Link to further Information

27.10.2011

Technical Specifications:

  • Measuring Method: Alkaline Haematin D Reagent Method
  • Measuring Range: 0.5-30g/dl
  • Display: 3 digits LED
  • Wavelength: 575 nm
  • Dilution: 1:101
  • Light Source: LED
  • Sample Volume: 2 ml
  • Detector: Silicon Photodiode
  • Accuracy: CV < ± 2%
  • Power Source: AC 220V/50Hz
  • Power Consumption: 15 W
  • Weight: 2 kg
  • Dimensions: 230(W) x 255(D) x 150(H)

 

Salient Features:

  • Compact design
  • Direct reading of Haemoglobin on digital display.
  • Long life 575 nm LED for accurate measurement.
  • Drain cell system designed to measure large number of samples.

Link to further Information

27.10.2011

Technical Specifications:

  • Assay Types: End-point with Sample Blank and Bichromatic, Multi-Standards, Fixed Time, Kinetic, Coagulation and Immunoturbidimetry.
  • Wavelengths: Interference filters 340, 405, 450, 505, 546, 578, 620 nm and two free l flow cell with Peltier element.
  • Temperature Control: 25, 30, 32 and 37ºC with precision of ±0.1ºC.
  • Aspiration Volume: From 0.5 ml to 1.0 ml in the steps of 0.1 ml for the flow cell.
  • Calibration: Automatic, manual & one point calibration before each sample
  • Cuvette Mode: Square cuvette of 10 mm path length to read 250µl. Micro cuvette for coagulation.
  • Storage Capacity: 100 programs, 999 test records.
  • Screen Size: 240 x 128 dots.
  • Display: Programs, processed data, curves for calibration and kinetic reactions, Levy-Jenning's curve for Quality control data, Online Temperature, Absorbance and light transmitting capacity in terms of Frequency.
  • Printer: Built-in Thermal Printer which can be switched ON/OFF, when desired.
  • Interface: RS 232C
  • Power supply: 230 V AC±10V, 50 Hz
  • External Dimensions: 9"(L) x 17"(W) x 20"(H).
  • Weight: 5 Kg

 

Salient Features:

  • Used for in vitro biochemical investigations.
  • Near ultraviolet and visible programmable with user friendly software.
  • Built-in clock and calendar displayed online.
  • Built-in 'TIME OUT' device enables the instrument to enter into the STANDBY mode when the instrument is not in use. In this mode intensity of the photometer lamp is reduced and backlit of LCD is switched off.
  • Self check program for filters, lamp and printer before display of main menu.
  • 6 digit 'PASSWORD' facility to program, modify or delete any procedure.
  • Zeroing, wavelength selection and temperature selection is automatic as per the chemistry programmed.
  • In end point and fixed time chemistries minimum 1 to maximum 10 calibrators can be used. 5 or more calibrators give a linearity curve which can be seen by a graphic display.
  • During run, online display of current temperature, absorbance, delay time, measuring time and frequency to check the performance of lamp and light transmitting capacity of the flow cell.
  • Kinetic reaction curves can be displayed, if desired.
  • 15 absorbance readings during kinetic run can be printed, if required, irrespective of measuring time given in the program and calculates change in absorbance per minute and correlation of the reaction.
  • Facility to prevent the carryover of previous sample by rinsing the flow cell, if and when required.
  • Result can be printed automatically, manually and also from the memory at any time until the stored data is erased.
  • Plots Levy & Jennings graph to monitor and provide statistics for quality control results.
  • All results of the patient with particular identification number are transferred to DATA PRINT in Utilities menu and can be displayed and printed out patient wise, if and when desired.

Link to further Information
27.10.2011

Technical Specifications:

  • Assay Types: End-point with Sample Blank and Bichromatic, Multi-Standard, Fixed Time, Kinetic, Non-linear, Coagulation and Immunoturbidimetry.
  • Wavelengths: Interference filters 340, 405, 450, 505, 546, 578, 620 nm and two free l flow cell with Peltier element.
  • Temperature Control: 25, 30, 32 and 37ºC with precision of ±0.1ºC.
  • Aspiration Volume: From 0.5 ml to 1.0 ml in the steps of 0.1 ml for the flow cell.
  • Cuvette Mode: Square cuvette of 10 mm path length to read 250µl. Micro cuvette for coagulation.
  • Storage Capacity: 200 programs, 999 test records.
  • Screen Size: 320 x 240 dots.
  • Display: Programs, processed data, Linear & Non-linear curves for calibration and kinetic reactions, Levy-Jenning's curve for Quality Control data.
  • Display: Online Temperature, Absorbance and light transmitting capacity in terms of Frequency.
  • Printer: Built-in Thermal Printer which can be switched ON/OFF, when desired.
  • Interface: RS 232C
  • Power supply: 110-230 V AC±10V, 50 Hz

 

Salient Features:

  • Used for in vitro biochemical investigations.
  • Near ultraviolet and visible programmable with user friendly software.
  • Built-in clock and calendar displayed online.
  • Built-in 'TIME OUT' device enables the instrument to enter into the STANDBY mode when the instrument is not in use. In this mode, the intensity of the photometer lamp is dimmed and LCD backlight is switched off.
  • Self check program for filters, lamp and printer before display of main menu.
  • 6 digit 'PASSWORD' facility to program, modify or delete any procedure.
  • Zeroing, wavelength selection and temperature selection are automatic as per the chemistry programmed.
  • In end point and fixed time chemistries minimum 1 to maximum 10 calibrators can be used. Five or more calibrators give a linearity curve that can be displayed, printed and stored in the memory.
  • Curve for non-linear calibration can be displayed, printed and stored in the memory.
  • In both linear as well as non-linear calibration, if any point is out of the curve it can be edited and corrected to calculate the corrected factor.
  • During run, online display provides current temperature, absorbance, delay time, measuring time and frequency to check the performance of lamp and light transmitting capacity of the flow cell.
  • Kinetic reaction curves can be displayed, printed and stored in the memory.
  • 15 absorbance readings during kinetic run can be printed, if required, irrespective of measuring time given in the program and change in absorbance per minute and correlation of the reaction can be calculated.
  • Automatic cleaning of flow cell is carried out before changing from one chemistry to another.
  • Facility to prevent the carryover of previous sample by rinsing the flow cell, if and when required.
  • Result can be printed automatically, manually and also from the memory at any time until the stored data is erased.
  • Levy & Jennings graph to monitor and provide statistics for quality control results can be displayed, printed and stored in the memory.
  • All results of the patient with particular identification number aretransferred to DATA PRINT in Utilities menu and can be displayed and printed out patient- wise, if and when desired.

Link to further Information